在地球化学样品测试中,目前对Ag、B、Sn的测定基本采用WP-Ⅰ型平面光栅摄谱仪、译谱仪读谱方法;其操作步骤复杂,在摄谱、读谱过程中受相板质量影响比较大,对读谱环境条件要求比较高。该文采用改装后的CCD-Ⅰ型平面光栅电弧直读发射光谱仪与WP-Ⅰ型平面光栅摄谱仪进行了对比,选用国家一级标准样品进行了Ag、B、Sn的检出限、精密度、准确度等实验,其结果表明:CCD-Ⅰ型平面光栅电弧直读发射光谱仪的测定结果优于WP-Ⅰ型平面光栅摄谱仪;该仪器采用CCD全谱单色器信息采集装置、中央控制器和数据采集功能,实现了摄谱后直接读取数据,完全摒弃了使用相板、洗相、译谱等过程;该方法具有分析速度快、分析结果精准、稳定性好的特点,可以一次曝光摄谱,满足1∶25万多目标区域地球化学调查规范要求。
In the analysis of geochemical samples,Ag,B and Sn are currently determined by WP-Ⅰplane grat-ing spectrograph and the reading is done by spectrum-interpreting device.The process is lengthy and the spectrograph and reading greatly depend on the quality of phase plate,requiring much from the reading environment.The paper compares renovated CCD-Ⅰlane grating electric arc direct reading emission spectrometer and WP-Ⅰplane grating spectrograph,and tests on the detection limit,accuracy and precision of Ag,B and Sn using first-grade national refer-ence substance.The results show that CCD-Ⅰlane grating electric arc direct reading emission spectrometer performs better than WP-Ⅰplane grating spectrograph;the device is equipped with CCD all-spectral monochromater collec-tion mechanism,central control unit and data collection system,and achieves direct data reading after spectrograph,to-tally giving up the process of phase plate,phase washing and spectrum interpretation;the device is good for its fast a-nalysis speed,accurate analysis results and stability.It can do spectrograph at once exposure and meet the requirement of 1∶250 000 multi-target regional geochemical survey.
参考文献
[1] | 张雪梅;张勤.发射光谱法测定勘查地球化学样品中银硼锡钼铅[J].岩矿测试,2006(4):323-326. |
[2] | 寇德凯;邱宏喜;冯俊;龙志武.发射光谱单电极载体蒸馏法测定地球化学样品中的银硼锡[J].黄金,2014(6):80-84. |
[3] | 俞晓峰;李锐;寿淼钧;郑存江;胡勇平.E5000型全谱直读型电弧发射光谱仪研制及其在地球化学样品分析中应用[J].岩矿测试,2015(1):40-47. |
[4] | 丁春霞;王琳;孙慧莹;刘艳霞;王筱丽.发射光谱法测定生态地球化学调查样品中的银锡硼[J].黄金,2012(10):55-58. |
[5] | 李林庆;杨菊;葛晓莹;戚洪友;焦圣兵;冯明辉;何丽;罗善霞.电弧蒸馏光谱法测定地球化学样品中银及其他痕量元素的研究进展[J].光谱实验室,2013(4):1947-1949. |
- 下载量()
- 访问量()
- 您的评分:
-
10%
-
20%
-
30%
-
40%
-
50%