欢迎登录材料期刊网

材料期刊网

高级检索

建立了小量水样痕量苯酚的液-液微萃取气相色谱分析法。考查了乙酸乙酯在酸和盐的作用下对苯酚的萃取性能,优化了萃取过程。采用FID检测器、WBI-17直接进样口和DB-1宽口径弹性石英毛细管柱,通过快速进样和汽化室内填装适当的玻璃棉可以得到满意的定量精密度。8 mL水样、160 μL乙酸乙酯、3.5 g硫酸铵可获得1 μg/L的检测下限,加标回收率为95.0%~98.5%,RSD为2.8%~3.3%。

A method for the liquid-liquid micro-extraction and gas chromatographic analysis of trace phenol in small amount of aqueous sample has been developed. The influences of acid and salts on recovery rate of phenol were examined. And, the influences of injection speeds and the conditions of silica wool in injection port on the quantitation precision were tested. With a direct injection port, an FID and a DB-1 capillary column of wide bore, a detection limit of 1 μg*L-1 can be obtained employing 8 mL water sample, 160 μL ethyl acetate and 3.5 g ammonium sulfate. The added recoveries were in the range of 95.0%-98.5%. The relative standard deviations were in the range of 2.8%-3.3%. This method is simple, convenient, rapid, accurate and practical.

参考文献

[1] Chinese Environment Protection Agency. Analysis meth-od for monitoring water and waste water.Third edition.Beijing:Environmental Science Press in China,1989.411-412国家环保局.水和废水监测分析方法.第3版. 北京:中国环境科学出版社,1989.411-412
[2] XU Ling, SUN Dong-hong. Environmental Monitoring in China,1997, 13:57-59徐泠,孙东红. 中国环境监测,1997, 13:57-59
[3] Realini P A. J Chromatogr Sci, 1981, 19:124-127
[4] Nick K, Scholer H F, Fresenius J. Anal Chem, 1992, 343:304-307
[5] Wennrich L, Efer J, Engewald W. J Chromatographia, 1995, 41:361-366
[6] GAO Zi-li. Solvent extraction chemistry. Beijing:Science Press,1991.6高自立. 溶剂萃取化学. 北京:科学出版社,1991.6