建立了固相微萃取(SPME)-高效液相色谱(HPLC)联用同时测定环境水样中8种多环芳烃的分析方法.优化了萃取时间、萃取温度、解吸时间、解吸溶液、解吸模式等条件.该法对8种多环芳烃的检出限为0.002~0.180 μg/L,相对标准偏差(RSD, n=6)为4.4%~12.2%.用该法分析江水中的痕量多环芳烃,除苯并[b]荧蒽外,其他7种多环芳烃的回收率为91.1%~115.8%,RSD(n=3)为3.6%~18.8%.方法快速、灵敏、简单,适用于快速分析环境水样中的痕量多环芳烃.
参考文献
[1] | Liu Shuqin, Wang Peng. Environmental Protection, 1995, (9):42刘淑琴, 王鹏. 环境保护, 1995, (9): 42 |
[2] | Potter D W, Pawliszyn J B. J Chromatogr, 1992, 625(1): 247 |
[3] | Chen J, Pawliszyn J B. Anal Chem, 1995, 67(9): 2 530 |
[4] | Wu Y C, Huang S D. Anal Chem, 1999, 71(2): 310 |
[5] | Jinno K, Muramatsu T, Saito Y, Kiso Y, Magdic S, Pawliszyn J. J Chromatogr A, 1996, 754: 137 |
[6] | Eisert R, Pawliszyn J. Anal Chem, 1997, 69(16): 3 140 |
[7] | Wu Y C, Huang S D. J Chromatogr A, 1999, 835: 127 |
[8] | Boyd-Boland A A, Pawliszyn J. Anal Chem, 1996, 68(9): 1 521 |
[9] | Negrao M R, Alpendurada M F. J Chromatogr A, 1998, 823: 211 |
[10] | Wang Chaoying, Tao Jingqi, Li Bifang, Ma Zhiling, Li Gongke. Chinese Journal of Chromatography, 2002, 20(1): 59王超英, 陶敬奇, 李碧芳, 马志玲, 李攻科. 色谱, 2002, 20(1): 59 |