本文介绍一种高复现性的新型低温固定点器件--带毛细管结构的4Heλ转变密封瓶的制作和使用方法.使用一般的绝热恒温器,只需简单控制平台和密封瓶下池的温度,即可方便地观测到不平度小于0.05mK长达数小时的4Heλ转变温坪.实验数据表明,通过毛细管的热流对λ转变温度测量值有一定的下压作用,其数值为100μW下压量为0.38mK.改变通过毛细管的热流,测出多个不同热流下的温坪值,外推到零热流,就可得到真实的Tλ温度(2.1768K).
参考文献
[1] | Comite Consultatif de Thermometrie (CCT) .Supplementary Information for the ITS-90 Bureau International des Poids et Mesures (BIPM)[Z].,1990. |
[2] | K F Hwang;B M Khorana .[J].Metrologia,1976,12:61-63. |
[3] | Song Nai-hao;Mao Yu-zhu;Lin Peng;Zhang Qing-geng,Zhang Liang.Proceedings of International Conference on Cryogenics and Refrigeration[M].Hangzhou China International Academic Publishers,1989:227. |
[4] | Nai-hao Song;C S Hong;Yu-zhu Mao;Peng Lin Qing-geng Zhang,Liang Zhang .[J].CRYOGENICS,1991,31:87-93. |
[5] | Lin Peng;Mao Yu-zhu;Hong Chao-sheng;Yue Yi,Zhang Qing-geng .[J].Cryogenics,1990,30:432. |
[6] | Robert V;Duncan;Guenter Ahlers .Temperature,Its Measurement and Control[J].Science and Industry,1992,6:243-245. |
[7] | Lin Peng;Mao Yu-zhu;Hong Chao-sheng.Proceedings of International Conference of the Temperature and Thermal Measuremeants[M].北京:中国标准出版社,1997:28. |
[8] | Guenter Ahlers .[J].Physical Review,1968,171:275-282. |