为获得具有表面多孔层的Ti6Al4V合金髋关节假体,选择不含Cu,Ni等元素的钎料,采用真空钎焊方法获得了纯钛/钛合金表面多孔层.钎焊工艺参数为950℃/30min时,采用φ0.2mm以上粒径的纯钛/钛合金球形粉末,孔径大于100μm,孔隙率6.0%~8.8%,表面多孔层与基体的结合强度大于20MPa.孔径、孔隙率与结合强度均满足人体内植入要求.所采用的钎焊热循环对Ti6Al4V合金股骨柄基体的力学性能和显微组织没有明显影响,符合标准YY 0117.1的规定;生物学评价遗传毒性实验结果为阴性,初步表明钎焊获得的纯钛/Ti6Al4V合金表面多孔层具有良好的生物相容性.

参考文献

[1] 吕厚山.人工关节外科学[M].北京:科学出版社,1998
[2] PILLIAR R M .Porous-surfaced metallic implants for orthopedic applications[J].Journal of Biomedical Materials Research,1987,21:1-33.
[3] THIEME M;WIETERS K P;BERGNER F et al.Titanium powder sintering for preparation of a porous functionally graded material destined for orthopaedic implants[J].Journal of Materials Science:Materials in Medicine,2001,12:225-231.
[4] LEE T;CHANG E;YANG C .Comparision of corrosion behavior and surface characteristics of vacuum brazed and heat treated Ti6Al4V alloy[J].Journal of Materials Science:Materials in Medicine,1998,9:429-437.
[5] YANG Y Z;TIAN J M;TIAN J T et al.Preparation of graded porous titanium coatings on titanium implant materials by plasma spraying[J].Journal of Biomedical Materials Research,2000,52:333-337.
[6] 周宇,杨贤金,崔振铎.新型医用β-钛合金的研究现状及发展趋势[J].金属热处理,2005(01):47-50.
[7] ASTM F 1147-2005.ASTM F 1147-2005.Standard test method for tension testing of calcium phosphate and metal coatings[S].
[8] ASTM F2068-2003.ASTM F2068-2003.Standard specification for femoral prostheses--metallic implants[S].
[9] YY 0117.1-2005.YY 0117.1-2005.外科植人物骨关节假体锻、铸件Ti6Al4V钛合金锻件[S].
[10] ISO 5832-3-1996.ISO 5832-3-1996.Implants of surgery-metallic materials-part 3:Wrought titanium 6-aluminium 4-vanadium alloy[S].
[11] GB/T 13810-2007.GB/T 13810-2007.外科植入物用钛及钛合金加工材[S].