提出了一种新型的CVD金刚石膜/硅(p~1010Ω·cm)为基板的微条气体室,阳极宽度7μm,阴极100 μm,间距200 μm.室温下,充以1.01×10-5Pa Ar+CH4混合气体时,采用5.9 keV 55FeX射线测量了探测器在不同高压和气体比例时的脉冲高度分布,详细讨论了高压和气体比例对能量分辨率的影响,结果表明该探测器具有较高的信噪比和能量分辨率.在Ar+10%CH4混合气体、-1100 V漂移电压和-650 V阴极电压工作条件下,微条气体室能量分辨率可达12.2%.
参考文献
[1] | Charpak G .Use of multiwire proportional counters to select and localize charged particles[J].Nuclear Instruments and Methods in Physics Research A:Accelerators,Spectrometers,Detectors and Associated Equipment,1968,62:262-264. |
[2] | Oed A .Position - sensitive detector with microstrip anode for electron multiplication with gases[J].Nuclear Instruments and Methods in Physics Research,1988,263:351-359. |
[3] | Bellazzini R;Bozzo M;Brez A et al.CMS micro -strip gas chamber project - development of a high - resolution tracking detector for harsh radiation environments[J].Nuclear Instruments and Methods in Physics Research A:Accelerators,Spectrometers,Detectors and Associated Equipment,2001,457:22-42. |
[4] | Fourme R. .POSITION-SENSITIVE GAS DETECTORS - MWPCS AND THEIR GIFTED DESCENDANTS[J].Nuclear Instruments and Methods in Physics Research, Section A. Accelerators, Spectrometers, Detectors and Associated Equipment,1997(1/3):1-11. |
[5] | D.S.A.P. Freitas;J.F.C.A. Veloso;C.M.B. Monteiro;J.M.F. dos Santos;C.A.N. Conde .The gas proportional scintillation counter/micro strip gas chamber hybrid detector[J].Nuclear Instruments and Methods in Physics Research, Section A. Accelerators, Spectrometers, Detectors and Associated Equipment,2003(1/2):228-232. |
[6] | Schulz Ch;Gebauer B;Richter G et al.Development of hybrid MSGC detectors with high position and time -of-flight resolution for neutron scattering experiments at ESS[J].Proceedings of SPIE-The International Society for Optical Engineering,2002,4785:203-213. |
[7] | Minglong Zhang;Yiben Xia;Linjun Wang;Weili Zhang .The electrical properties of diamond-like carbon film/D263 glass composite for the substrate of micro-strip gas chamber[J].Diamond and Related Materials,2003(9):1544-1547. |
[8] | Oed A .Properties of micro-strip gas chambers (MSGC)and recent developments[J].Nuclear Instruments and Methods in Physics Research A:Accelerators,Spectrometers,Detectors and Associated Equipment,1995,367:34-40. |
[9] | Bateman JE.;Connolly JF.;Derbyshire GE.;Duxbury DM.;Mir JA.;Spill EJ. Stephenson R. .Energy resolution in X-ray detecting micro-strip gas counters[J].Nuclear Instruments and Methods in Physics Research, Section A. Accelerators, Spectrometers, Detectors and Associated Equipment,2002(1/3):384-395. |