采用低温辐射计在1064 nm波长对两个硅陷阱探测器(trap O#和trap B)进行绝对光功率响应度定标.描述了定标方法、实验装置和数据的分析方法,提出了低温辐射计窗口的精确复位和窗口透过率的测量方案,测量得到的透过率达到0.999以上.结果表明,两个探测器绝对响应度定标的合成不确定度分别为1.457×10-4和1.458×10-4,定标结果的重复性分别达到了0.014%和0.008%.分析了测量时各项不确定度来源,对光功率响应度定标不确定度进行了评估.对O#硅陷阱探测器在1064 nm波长处光功率响应度的温度稳定性进行了测试研究.
参考文献
[1] | Fox N P.Trap detectors and their properties[J].Metrologia,1991,28:197-202. |
[2] | Gardner J L.Transmission trap detectors[J].Appl.Opt.,1994,33(25):5914-5918. |
[3] | Zheng X B,Wu H Y,Zhang J P.Absolute spectral responsivity standard detectors with uncertainty less than 0.035%[J].Acta Optica Sinica,2001,21(6):749-752 (in Chinese). |
[4] | 5zcan Bazklr,Sevilay Ugur,Farhad Samedov.High-accuracy optical power measurements by using electricalsubstitution cryogenic radiometer[J].Opt.Eng.2005,44(1):016401-1-016401-6. |
[5] | Li Shuang,Wu Haoyu,Li Zhaozhou,et al.The experiment study of high accuracy optical radiation standard detectors in NIR[J].Optical Technology (光学技术),2004,30(4):498-501 (in Chinese). |
[6] | Stefan Kiick,Helmuth Hofer,Marco Antonio LSpez Ordonez.Cryogenic radiometer-based high accurate measurement of Ge and InGaAs trap detector responsivity[J].Appl.Opt.,2006,45(3). |
[7] | Hamamatsu,Photodiode technical information,http://sales.hamamatsu.com/en/products-type.php. |
[8] | Li Shuang.High accurate optical radiometric calibration based on cryogenic radiometer[D].Hefei:Master Thesis of Anhui Institute of Optics and Fine Mechanics (中国科学院安徽光机所硕士论文),CAS,2003.(in Chinese). |
[9] | Li Zhaohou,Zheng X B,Wu Haoyu.Temperature characteristic study of high accuracy optical radiometric standard detectors[J].Acta Optica Sinica (光学学报),2004,24(3):401-407 (in Chinese). |
上一张
下一张
上一张
下一张
计量
- 下载量()
- 访问量()
文章评分
- 您的评分:
-
10%
-
20%
-
30%
-
40%
-
50%