利用Pro/E和Ansys软件建立了胸主动脉支架轴向柔顺性分析模型,对27种NiTi合金支架的柔顺性进行了模拟分析与实验验证。在此基础上,又进一步系统地分析了弯曲变形时支架结构尺寸变化对中央部位横截面变形行为的影响。结果显示,减少圆周方向上的连接体的数量,增加支撑体的长度或数量,都能够提高支架的柔顺性,但是减少连接体的数量比增加支撑体的长度或数量对支架柔顺性的影响程度要大。支架弯曲变形时连接体的数量对支架中央部位横截面的变形具有显著的影响。

The models of chest aorta stent for axial flexibility analysis were established by software of Pro/E and Ansys. The flexibility of twenty-seven NiTi alloy stent was simulated and verified. On this basis, the effect to performance of cross section in central part was analyzed systematically when stent structure was changed. The results showed that flexibility of stent could been improved by decreasing linkers of circumferential direction as well as increasing length or number of cell, but the former is more significant than the latter. Cross section de- formation in central part was affected obviously by the number of linkers when stent structure was changed.

参考文献

[1] 符伟国,董智慧.主动脉瘤和主动脉夹层腔内治疗的问题与对策[J].外科理论与实践,2005(01):15-18.
[2] 杨建勇;陈伟.介入放射学理论与实践[M].北京:科学出版社,2005:112-113.
[3] Von Kodolitsch Y;Csosz S K;Koschyk D H et al.Intramural hematoma of the aorta:predictors of progres- sion to dissection and rupture[J].Circulation,2003,107(08):1158-1163.
[4] Marvin D;James H;Richard P .Aortic dissection,per- spectives in the era of stent-graft repair[J].Journal of Vascular Surgery,2006,43(02):30.
[5] Szmid T J;Rowi S O;Ga Zka Z .Simultaneous endovasc- ullar exclusion of thoracic aortic aneurysm with open ab- dominal aortic aneurysm repair[J].Euro J Vas Endo Vasc Surg,2004,28(04):442.
[6] Migliavacca F;Petrini L;Colombo M;Auricchio F;Pietrabissa R .Mechanical behavior of coronary stents investigated through the finite element method.[J].Journal of Biomechanics,2002(6):803-811.
[7] Gong Xiaoyan;Alan R P.ABAQUS analysis on nitinol medical applieations[A].Newport,2002
[8] Whitcher F D .Simulation of in vivo loading conditions of Nitinol vascular stent structures[J].Computers & Struc- tures,1997,64(5-6):10051011.
[9] Petrini L;Migliavaacca F;Dubini G.Numerical analysis of vascular stents exploiting shape-memory-alloy behavior[A].Meccanica,2003:1-10.
[10] Henry M;Klonafis C;Amor M et al.State of the art:which stent for which lesion in peripheral interventions[J].Texas Heart Institute Journal,2000,27(02):119-126.
[11] Raval A;Choube B;Eneineer C et al.Development and assessment of 316LMV cardiovascular stents[J].Materials Science and Engineering,2004,386(1-2):331-343.
[12] Ormiston JA;Dixon SR;Webster MW;Ruygrok PN;Stewart JT;Minchington I;West T .Stent longitudinal flexibility: a comparison of 13 stent designs before and after balloon expansion.[J].Catheterization and cardiovascular interventions: Official journal of the Society for Cardiac Angiography & Interventions,2000(1):120-124.
[13] Petrini L.;Migliavacca F.;Auricchio F.;Dubini G. .Numerical investigation of the intravascular coronary stent flexibility[J].Journal of Biomechanics,2004(4):495-501.
[14] 扈玉玲,缪卫东,许尚栋,郭锦芳.新型TiNi合金胸主动脉支架的研制与性能研究[J].稀有金属,2008(05):584-588.
[15] 付文宇 .基于逆向工程的动脉瘤及支架的数值模拟研究[D].北京工业大学,2010.
[16] 梁栋科 .血管内支架的加工及其力学性能的分析与评价[D].大连理工大学,2005.
[17] 柳清菊,张瑾,朱忠其,张海滨,吴兴惠.载银TiO2无机抗菌剂的制备及性能研究[J].功能材料,2005(03):474-476.
[18] 高振宇,梁栋科,齐民,杨大智,王伟强.镍钛合金超弹性支架抗压缩性能的实验与分析[J].功能材料与器件学报,2006(01):49-53.
[19] 高振宇,梁栋科,齐民,张庆宝,杨大智.镍钛合金超弹性支架纵向柔顺性分析[J].功能材料,2007(01):113-114.
[20] 张庆宝,王伟强,齐民,梁栋科,杨大智.不同材料冠状动脉支架膨胀行为分析[J].功能材料,2007(01):130-134.
[21] 刘强,程晓农,徐红星,费黄霞,许文荣.316L不锈钢和NiTi合金血管支架的血液相容性[J].中国组织工程研究与临床康复,2008(04):735-738.
[22] Gong X Y;Pelton A R.Proceedings of the international conference on shape memory and superelastic technolo- gies[M].USA:Fremont,2003:443-451.
[23] Takashima K;Kitou T;Mori K;Ikeuchi K .Simulation and experimental observation of contact conditions between stents and artery models.[J].Medical engineering and physics,2007(3):326-335.
上一张 下一张
上一张 下一张
计量
  • 下载量()
  • 访问量()
文章评分
  • 您的评分:
  • 1
    0%
  • 2
    0%
  • 3
    0%
  • 4
    0%
  • 5
    0%