文章介绍了MTS810.25T疲劳试验机的工作站与计算机的无法接通故障的情况、分析出现该故障的原因以及制定相应的维修方案。本次故障是电源板上的一个电容损坏所造成的,更换电容之后,试验机恢复正常使用。
The disconnection between computer and workstation of MTS810.25T fatigue testing machine is described. Meanwhile, the breakdown reasons are analyzed and repair plan are set down. By replacing a damaged capacitor with good one, the testing machine is got right.