欢迎登录材料期刊网

材料期刊网

高级检索

以戊二酸锌为催化剂合成可降解聚碳酸亚丙亚乙酯(PPEC,MW=229.338kDa)并首次将其作为载药涂层材料进行研究。核磁共振氢谱(1 H NMR)显示分子链中环氧乙烷和环氧丙烷的贡献与原料中两者配比相同,碳酸酯段含量为76.7%。差示扫描量热仪(DSC)测得玻璃化转变温度(Tg)为16.8℃。拉伸实验得到断裂伸长率为550%。衰减全反射-傅立叶变换红外光谱(ATR-FT-IR)显示药物涂层中PPEC和雷帕霉素之间没有发生明显的化学反应。球囊扩张实验后用扫描电子显微镜(SEM)观察到药物涂层支架表面完整光滑,没有剥落翘起的现象。药物洗脱支架在模拟体液(PBS,pH值=7.4)中药物释放时间超过60d,速率由快变慢。实验表明PPEC作为支架涂层材料具有广阔的应用前景。

The biodegradable poly(propylene ethylene carbonate)(PPEC,M w =229.338kDa) was synthesized using zinc glutarate catalyst and investigated as drug-loaded coating materials for the first time in this paper.1 H NMR spectra show the contribution of ethylene oxide and propylene oxide is 6∶4and the carbonate content is 76.7%in PPEC molecular formula.The elongation at break is 550%by stress-strain test.Attenuated total reflectance-infrared spectroscopy(ATR-FT-IR) analysis of sirolimus,PPEC and sirolimus-loaded PPEC coating show that no chemical reactions happene between sirolimus and PPEC.After balloon expansion experiment,no delamination or destruction of the coating is observed by SEM.In vitro drug release behavior reveals that drug release of drug-eluting stent is over 60days in PBS(0.05mol / L,pH =7.4) solution.All the experiment results show that PPEC as a stent coating materials has broad application prospects.

参考文献

[1] Stone GW;Moses JW;Ellis SG;Schofer J;Dawkins KD;Morice MC;Colombo A;Schampaert E;Grube E;Kirtane AJ;Cutlip DE;Fahy M;Pocock SJ;Mehran R;Leon MB .Safety and efficacy of sirolimus- and paclitaxel-eluting coronary stents.[J].The New England journal of medicine,2007(10):998-1008.
[2] Sigwart U .[J].Journal of the American College of Cardiology,2006,47:1361-1362.
[3] Stettler C;Wandel S;Allemann S;Kastrati A;Morice MC;Schomig A;Pfisterer ME;Stone GW;Leon MB;de-Lezo JS;Goy JJ;Park SJ;Sabate M;Suttorp MJ;Kelbaek H;Spaulding C;Menichelli M;Vermeersch P;Dirksen MT;Cervinka P;Petronio AS;Nordmann AJ .Outcomes associated with drug-eluting and bare-metal stents: a collaborative network meta-analysis.[J].The Lancet,2007(9591):937-948.
[4] Bavry A A;Kumbhani D J;Helton T J et al.[J].American Journal of Medicine,2006,119:1056-1061.
[5] Rainer W .[J].Nature Reviews Cardiology,2010,7:194-203.
[6] Wang XT;Venkatraman SS;Boey FYC;Loo JSC;Tan LP .Controlled release of sirolimus from a multilayered PLGA stent matrix[J].Biomaterials,2006(32):5588-5595.
[7] 黄莹莹,齐民,杨大智,王昌明.冠脉支架表面PLGA涂层制备及其血液相容性研究[J].功能材料,2006(03):411-414.
[8] Han Yaling;Jing Quanmin;Li Yi et al.[J].Catheterization and Cardiovascular Interventions,2012,79(02):211-216.
[9] Byrne RA;Kastrati A;Massberg S;Wieczorek A;Laugwitz KL;Hadamitzky M;Schulz S;Pache J;Fusaro M;Hausleiter J;Schomig A;Mehilli J .Biodegradable polymer versus permanent polymer drug-eluting stents and everolimus- versus sirolimus-eluting stents in patients with coronary artery disease: 3-year outcomes from a randomized clinical trial.[J].Journal of the American College of Cardiology,2011(13):1325-1331.
[10] Holmes Jr;Kereiakes D J;Garg S et al.[J].Journal of the American College of Cardiology,2010,56:1357-1365.
[11] Salinas P;Moreno R;Jimenez Valero S;Galeote G;Sanchez Recalde A;Plaza I;Lopez Sendon J .Stent thrombosis did biodegradable polymers fail or are we too impatient?[J].Journal of the American College of Cardiology,2011(8):885-886.
[12] James K;Kohn J .[J].Materials Research Society Bulletin,1996,21(11):22-26.
[13] EDLUND ULRICA;ALBERTSSON ANN-CHRISTINE .Copolymerization and Polymer Blending of Trimethylene Carobonate and Adipic Anhydride for Tailored Drug Delivery[J].Journal of Applied Polymer Science,1999(2):227-239.
[14] Kawakuchi T;Nakano M;Juni K et al.[J].Chemical and Pharmaceutical Bulletin,1983,31(04):1400-1403.
[15] Acemoglu M .Chemistry of polymer biodegradation and implications on parenteral drug delivery.[J].International Journal of Pharmaceutics,2004(1/2):133-139.
[16] Zhou M;Takayanagi M;Yoshida Y et al.[J].Polymer Bulletin,1999,42:419-424.
[17] Zhong X;Dehghani F .[J].Applied Catalysis B:Environmental,2010,98:101-111.
[18] 闵加栋,张亚明,陈玉岩,王献红,赵晓江,王佛松.稀土三元催化剂对二氧化碳/环氧化合物共聚反应的催化机理研究[J].高分子学报,2009(03):233-237.
[19] Antony P;Puskas J E;Kontopoulou M .[J].Polymer Engineering &Science,2003,43(01):243-253.
[20] Toutouzas K;Colombo A;Stefanadis C .[J].European Heart Journal,2004,25(19):1679-1687.
[21] 王伟强,梁栋科,杨大智,齐民.冠脉支架系统瞬时膨胀过程的有限元分析及其优化设计[J].中国生物医学工程学报,2005(03):313-318,329.
[22] 谭丽丽,薛松,杨柯,张炳春,刘冬雪,梁勇.冠脉支架表面载药涂层的制备和性能[J].材料研究学报,2004(04):380-384.
[23] Finkelstein A;McClean D;Kar S et al.[J].Circulation,2003,107:777-784.
上一张 下一张
上一张 下一张
计量
  • 下载量()
  • 访问量()
文章评分
  • 您的评分:
  • 1
    0%
  • 2
    0%
  • 3
    0%
  • 4
    0%
  • 5
    0%