采用脉冲衰减法可以测量生物组织热导率.但是,Chen等人提出的脉冲衰减法采用了"点热源"假设,而实际测量时,探头有一定的几何尺寸,由此造成了测量时的模型误差.本文采用数值实验的方法对这一误差及其影响因素进行了分析.在此基础上,提出了一种考虑了探头尺寸影响的新的脉冲衰减法,推导出了其测量公式.采用已知热导率的工质进行了验证试验,实验测量结果证实了本文对原有测量方法误差的分析;同时也表明,采用本文提出方法可以将测量误差减小到5%以下.
参考文献
[1] | Valvano J W;Allen J T;Bowman H F .The Simultaneous Measurement of Thermal Conductivity, Thermal Diffusivity, and perfusion in Small Volumes of Tissue[J].Journal of Biomechanical Engineering,1984,106(04):192-197. |
[2] | Chen M M;Holmes K R;Rupinskas V .Pulse-Decay Method for Measuring the Thermal Conductivity of Living Tissue[J].Journal of Biomechanical Engineering,1981,103(06):253-260. |
[3] | Arkin H;Holmes K R;Chen M M .A Sensitivity Analysis of the Thermal Pulse Decay Method for Measurement of Local Tissue Conductivity and Blood Perfusion[J].Journal of Biomechanical Engineering,1986,108(01):54-58. |
[4] | Diederich C J;Clegg S;Roemer R B .A Spherical Source Model for the Thermal Pulse Decay Method of Measuring Blood Perfusion: A sensitivity Analysis[J].Journal of Biomechanical Engineering,1989,111(01):55-61. |
[5] | Patel P A;Valvano J W;Pearce J A et al.A SelfHeated Thermistor Technique to Measure Effective Thermal Properties from the Tissue Surface[J].Journal of Biomechanical Engineering,1987,109(05):330-335. |