To investigate the configurations of the valence protons in Borromean nucleus 17Ne, a CsI(Tl) detector array, which consists of 9 CsI crystals (26×26×20mm3) coupled with photodiodes, has been successfully used in the 17Ne experiment to measure the energy of protons. In order to find the optimal working conditions and get the best energy resolutions, several technologies (including various wrapping materials, wrapping and coupling methods) have been used. The testing results showed that the best energy resolution of the CsI(Tl) is about 3.3% using the 241 Am α-source. The primary testing results with the proton beam were also provided.

参考文献

[1] Birks J B.The Theory and Practice of Scintillation Counting[A].
[2] Kubota Y et al.[J].Nuclear Instruments and Methods in Physics Research A:Accelerators,Spectrometers,Detectors and Associated Equipment,1992,320:66-113.
[3] Abashian A.;Abe K.;Abe R.;Adachi I.;Ahn BS.;Ahn HS.;Aihara H. Akatsu M.;Akhmetshin R.;Alexander JP.;Alimonti G.;An Q.;Aoki K.;Asai M.;Asano Y.;Aso T.;Aulchenko V.;Bakich AM.;Banas E.;Behari S. Behera PK.;Beiline D.;Bondar A.;Bozek A.;Browder TE.;Takasaki F.;et al. .The Belle detector[J].Nuclear Instruments and Methods in Physics Research, Section A. Accelerators, Spectrometers, Detectors and Associated Equipment,2002(1):117-232.