电感耦合等离子体质谱(ICP-MS)法具有很多优点,但其测定镀金液中杂质元素时基体元素Au对测定有干扰和抑制.用甲基异丁基酮(MIBK)萃取分离镀金液中基体Au,通过加入内标混合溶液校正残余Au对待测元素的干扰,再用ICP-MS测定镀金液中的杂质元素Mn,Fe,Cu,Co,Ni,Cd,Pb,讨论了萃取条件、仪器工作参数、同位素及内标元素、质谱干扰、基体效应等对测量的影响,确定了最佳测定条件.结果表明:该方法的检出限为0.015~0.180μg/L,加标回收率为93.22%~107.78%,相对标准偏差为1.6%~4.1%;该方法准确、快速、简便,完全满足镀金液中杂质元素的测定要求.
参考文献
[1] | 王玉宝,郭振良,唐清华,张江,孙琳,王向红.甲基紫褪色光度法测定酸性镀金液中微量的钴[J].电镀与涂饰,2003(05):49-50. |
[2] | 沈卓身,杨晓战,魏忻,任菲.镀金液中铜杂质的分光光度分析[J].材料保护,1999(10):17. |
[3] | Ivanova E;Havesov I;Berndt H et al.ET-AAS determination of traces of As,Sb and Sn in pure gold after matrix separation with hydrazine[J].Fresenius'Journal of Analytical Chemistry,1990,336(04):320-321. |
[4] | 何晓梅.基体分离-ICP-AES法测定镀金液中铅、铜、铁、镍的研究[J].冶金分析,2005(01):39-41. |
[5] | Graham S M;Robert R V D .The analysis of high-purity noble metals and their salts by ICP-MS[J].Talanta,1994,41(08):1369-1375. |
[6] | 孙丹丹,曹昕宇.萃取分离ICP-AES法测定高纯金中的杂质元素[J].光谱学与光谱分析,2001(06):849-851. |
[7] | Olivares J A;Houk R S .Suppression of analyzed signal by various concomitant salts in inductively coupled plasma mass spectrometry[J].Analytical Chemistry,1986,58(01):20-25. |